Skip to main content

Table 2 Prediction of validation set samples using the proposed chemometric models

From: Application of smart chemometric models for spectra resolution and determination of challenging multi-action quaternary mixture: statistical comparison with greenness assessment

Sample No.

Concentration (Āµg mLāˆ’ā€‰1)

PCR

PLS

%Recovery

%Recovery

PARA

ASC

CAF

CPM

PARA

ASC

CAF

CPM

PARA

ASC

CAF

CPM

26

20.00

15.00

2.50

1.00

98.88

98.96

101.91

100.96

98.88

96.93

101.87

101.07

27

4.00

15.00

2.50

9.00

102.54

98.72

98.96

98.37

102.44

98.39

98.92

98.42

28

8.00

8.00

8.00

8.00

99.91

98.83

99.53

100.48

100.01

99.25

99.54

100.53

29

12.00

9.00

7.50

5.00

99.40

101.87

96.16

99.87

99.44

101.47

96.16

99.92

30

20.00

3.00

12.50

1.00

99.81

100.61

96.06

100.08

99.86

101.49

96.07

100.51

Mean

100.11

99.80

98.53

99.59

100.12

99.51

98.51

100.09

SD

1.42

1.39

2.47

0.749

1.36

1.99

2.45

1.02

Ā 

MCR-ALS

ANN

26

20.00

15.00

2.50

1.00

100.87

101.02

98.95

102.64

99.83

100.00

100.01

100.00

27

4.00

15.00

2.50

9.00

102.62

98.70

96.49

96.51

100.00

100.00

100.00

100.00

28

8.00

8.00

8.00

8.00

98.08

101.97

98.02

99.35

100.00

100.00

100.00

104.52

29

12.00

9.00

7.50

5.00

99.74

101.57

96.13

100.10

101.16

99.52

101.95

100.00

30

20.00

3.00

12.50

1.00

100.48

101.57

99.43

100.64

100.00

100.00

100.00

100.00

Mean

100.29

99.88

97.80

99.85

100.20

99.88

100.49

100.63

SD

0.58

0.24

1.46

2.23

0.54

0.24

0.98

1.26