Skip to main content

Table 3 Effect of wavelength ranges on the recovery results of the prepared mixtures of CI and OR

From: The effectiveness of multivariate and univariate spectrophotometric techniques for the concurrent estimation of ornidazole and ciprofloxacin HCl in tablet formulation and spiked serum: estimating greenness and whiteness profile

Wavelength range

Linear equation

Recovery% Drugs concentration (μg/ml) CI: OR

10:5

5:15

3:10

CI

240–290 nma

Y = 0.2091X + 0.0026

99.42

98.92

98.10

230–300 nm

Y = 0.2630X + 0.0030

95.70

96.61

96.41

260–290 nm

Y = 0.1206X + 0.0015

96.07

95.18

92.98

240–325 nm

Y = 0.2943X + 0.0037

95.96

92.08

92.14

250–300 nm

Y = 0.2118X + 0.0028

97.03

95.78

96.82

OR

260–340 nma

Y = 0.0571X + 0.0042

98.13

99.12

101.09

290–330 nm

Y = 0.0231X + 0.0030

96.27

93.33

97.06

240–340 nm

Y = 0.0609X + 0.0040

104.21

103.94

104.80

260–330 nm

Y = 0.0581X + 0.0041

97.02

97.15

103.06

280–340 nm

Y = 0.0401X + 0.0037

98.37

93.71

97.03

  1. aThe elected wavelength ranges for mean centering transforms