Skip to main content

Table 5 Statistical comparison of the results obtained by the proposed TLC-densitometry method and the reported HPLC method for the analysis of Capozide ® tablet

From: AQbD TLC-densitometric method approach along with green fingerprint and whiteness assessment for quantifying two combined antihypertensive agents and their impurities

Method

TLC-densitometry

Reported methodb [10]

Parameters

Captopril

 Mean (%)a

101.96

102.08

 SD

0.34

0.37

 N

6

6

 Student’s t test (2.228)c

0.59

–

 F test (5.505)c

1.18

–

Parameters

Taken (µg/mL)

Recovery %

Standard addition technique

 

1.20

100.79

1.50

100.85

2.00

100.55

 Mean ± SD

100.73 ± 0.16

Parameters

  

Hydrochlorothiazide

  

 Mean (%)a

101.55

101.30

 SD

0.40

0.30

 N

6

6

 Student’s t test (2.228)c

1.26

–

 F test (5.505)c

1.78

–

Parameters

Taken (µg/mL)

Recovery %

Standard addition technique

 

0.50

100.29

0.70

100.35

1.00

100.56

 Mean ± SD

100.40 ± 0.14

  1. aAverage of percentage recoveries of six determinations
  2. bReported method is HPLC for determination of CPL and HCZ using methanol: water (45:55, v/v), as the mobile phase with UV detection at 210 nm
  3. cFigures between parentheses represent the corresponding tabulated values of t and F at p = 0.05