Skip to main content

Table 2 Accuracy evaluation of the proposed first derivative synchronous spectrofluorometric method using standard addition method on laboratory-made tablet formulation

From: Development of a green synchronous spectrofluorimetric technique for simultaneous determination of Montelukast sodium and Bilastine in pharmaceutical formulations

Amount taken (ng mL−1)

Amount added (ng mL−1)

Amount found (ng mL−1)

% Recovery ± SDa

MTK

BIL

MTK

BIL

MTK

BIL

MTK

BIL

50

100

0

0

50.24

99.89

100.47 ± 0.64

99.90 ± 0.83

50

100

150

100

199.04

200.39

99.52 ± 1.28

100.19 ± 1.32

50

100

750

300

801.59

398.75

100.20 ± 1.31

99.69 ± 1.77

50

100

1350

500

1409.25

604.06

100.66 ± 0.73

100.68 ± 0.39

  1. aMean of five determinations