Skip to main content

Table 1 Overview of the methodology validation parameters of the gallic acid and bergenin

From: Microwave extraction and molecular imprinted polymer isolation of bergenin applied to the dendrochronological chemical study of Peltophorum dubium

Parameter

Gallic acid

Bergenin

Linearity (R2)

0.9998

0.9996

Precision (%)

4.32

3.50

Accuracy (%)

98.83%–101.63

89.88%–102.10

LoD

0.12 µg/mL

0.01 µg/mL

LoQ

0.38 µg/mL

0.05 µg/mL

Calibration curves

y = 26114x − 18,484

y = 8400.9x − 35,785