Skip to main content

Table 4 Working conditions for measurements by HR-CS FAAS ContrAA 300 in air-acetylene flame

From: Validation of an analytical method based on the high-resolution continuum source flame atomic absorption spectrometry for the fast-sequential determination of several hazardous/priority hazardous metals in soil

Analyte

Wavelength (nm)

Acetylene (l/min)

Air (l/min)

Burner height (mm)

Ag

328.068

50

500

6

Cd

228.802

50

500

6

Co

240.206

50

500

6

Cr

357.869

100

400

8

Cu

324.754

50

470

6

Ni

232.003

55

470

6

Pb

217.000

65

470

6

Zn

213.856

50

470

6