Skip to main content

Table 3 Assay validation sheet of the proposed methods for the simultaneous determination of AML, OLM and HCT

From: Application of smart spectrophotometric methods and artificial neural network for the simultaneous quantitation of olmesartan medoxamil, amlodipine besylate and hydrochlorothiazide in their combined pharmaceutical dosage form

Parameter

Spectrophotometric methods

ANN method

Direct

Isoabsorptive

Ratio subtraction

AML

OLM

HCT

AML

OLM

HCT

Accuracy (mean ± RSD)

99.89 ± 0.937

101.96 ± 0.622

99.55 ± 1.889

100.47 ± 1.197

99.9 ± 1.066

100.75 ± 2.072

Repeatabilitya

0.552

0.829

0.587

1.321

1.061

1.144

Intermediate precisionb

0.925

1.327

1.593

1.119

1.683

2.424

Linear range (μg mL-1)

5-40

2.5-40

5-40

3-40

2.5-40

5-40

Slope

0.011

0.041

0.010

1.012

0.976

0.998

Intercept

0.003

0.077

0.01

−0.035

0.390

0.041

Correlation coefficient (r)

0.9999

0.9989

0.9999

0.9992

0.9995

0.9995

LOD (μg mL-1)

1.278

0.729

0.819

0.291

0.683

0.444

LOQ (μg mL-1)

4.217

2.209

2.482

0.880

2.070

1.345

  1. a The intraday (n = 3), average of concentrations; 6.95,20,12.5 μg mL-1for AML, OLM and HCT respectively, repeated three times within the day.