Skip to main content

Table 5 Analytical Characteristics for analysis of PA, MP, PP, CH and PS by ILS method

From: Resolution of five-component mixture using mean centering ratio and inverse least squares chemometrics

Analyte

Calibration equations

R2

Linear concentration range (μg mL-1)

LOD (μg mL-1)

PA

Eq. (11)

0.9999

0-25.6

0.08

MP

Eq. (12)

0.9992

0-15.0

0.04

PP

Eq. (13)

0.9985

0-15.0

0.03

CH

Eq. (14)

0.9981

0-45.0

0.05

PS

Eq. (15)

0.9971

0-100.0

0.06