Table 7 Accuracy and precision (within and between-run) of the six drugs in spiked human plasma

From: Therapeutic drug monitoring of six contraindicated/co-administered drugs by simple and green RP-HPLC-PDA; application to spiked human plasma

Drugs

Concentrations in plasma (µg/mL)

Within run (n = 3)

Between run (n = 12)

Mean accuracy (%R*)

Precision (%RSD**)

Mean accuracy (%R)

Precision (%RSD)

IMP

LLOQ

10.0

99.73

1.22

98.56

1.55

QCL

30.0

99.03

0.68

98.45

1.08

QCM

40.0

100.37

0.21

99.52

1.19

QCH

75.0

99.94

0.77

99.24

0.66

ETP

LLOQ

15.0

99.09

0.69

98.75

0.33

QCL

50.0

100.83

0.44

99.43

1.17

QCM

140.0

99.79

0.28

98.64

0.99

QCH

280.0

99.92

0.64

99.08

0.92

CLS

LLOQ

10.0

100.52

0.33

99.05

0.96

QCL

30.0

99.35

0.86

98.63

0.81

QCM

40.0

99.45

0.50

98.95

1.08

QCH

75.0

100.28

0.55

99.15

0.93

MRP

LLOQ

5.0

99.19

1.88

98.37

0.22

QCL

15.0

99.60

0.70

97.91

0.99

QCM

35.0

99.69

0.68

99.17

0.44

QCH

75.0

100.50

0.46

98.84

1.25

PRB

LLOQ

5.0

99.32

0.81

98.32

0.76

QCL

15.0

99.72

0.40

98.27

1.37

QCM

35.0

99.90

0.65

98.17

1.62

QCH

75.0

100.04

0.60

98.35

1.50

WRF

LLOQ

1.0

98.56

0.19

97.72

0.63

QCL

6.0

99.61

0.97

98.15

1.23

QCM

12.0

98.14

0.31

97.26

0.83

QCH

25.0

99.72

0.78

99.18

0.66

  1. *%Recovery