Potentiometric determination of moxifloxacin in some pharmaceutical formulation using PVC membrane sensors

Abstract

Background

The construction and electrochemical response characteristics of Poly (vinyl chloride) membrane sensors for moxifloxacin HCl (MOX) are described. The sensing membranes incorporate ion association complexes of moxifloxacin cation and sodium tetraphenyl borate (NaTPB) (sensor 1), phosphomolybdic acid (PMA) (sensor 2) or phosphotungstic acid (PTA) (sensor 3) as electroactive materials.

Results

The sensors display a fast, stable and near-Nernstian response over a relative wide moxifloxacin concentration range (1 × 10-2 - 4.0 × 10-6, 1 × 10-2 - 5.0 × 10-6, 1 × 10-2 - 5.0 × 10-6 M), with detection limits of 3 × 10-6, 4 × 10-6 and 4.0 × 10-6 M for sensor 1, 2 and 3, respectively over a pH range of 6.0 - 9.0. The sensors show good discrimination of moxifloxacin from several inorganic and organic compounds. The direct determination of 400 μg/ml of moxifloxacin show an average recovery of 98.5, 99.1 and 98.6% and a mean relative standard deviation of 1.8, 1.6 and 1.8% for sensors 1, 2 and 3 respectively.

Conclusions

The proposed sensors have been applied for direct determination of moxifloxacin in some pharmaceutical preparations. The results obtained by determination of moxifloxacin in tablets using the proposed sensors are comparable favorably with those obtained using the US Pharmacopeia method. The sensors have been used as indicator electrodes for potentiometric titration of moxifloxacin.

Background

Moxifloxacin (MOX) (1-cyclopropyl-6-fluoro-1,4- dihydro-8-methoxy-7-[(4aS,7aS)-octahydro-6H-pyrrolo [3,4-b]pyridin-6-yl]4-oxo-3 quinoline carboxylic acid) (Figure 1) is an advanced-generation, 8-methoxyquinolone derivate of fluoroquinolone antibacterial agent that is synthetic. It was discovered in 1999 [1],[2]. Moxifloxacin is a broad-spectrum antibiotic that is active against both Gram-positive and Gram-negative bacteria. It functions by inhibiting DNA gyrase, a type II topoisomerases, and topoisomerase IV [3], enzymes necessary to separate bacterial DNA, thereby inhibiting cell replication.

Figure 1
figure 1

Chemical structure of moxifloxacin hydrochloride.

Various methods cited in literature for its determinations involve, spectrophotometry [4],[5], spectrofluorimetry [6], atomic absorption spectrometry [7], conductometry [7], voltammetry [8], high performance liquid chromatography- ultraviolet (HPLC-UV) [9]-[11], HPLC-fluorescence (HPLC-Fl) [12]-[14] capillary electrophoresis (CE) [15],[16], and HPLC-mass spectrometry (HPLC-MS) [17],[18]. However, most of these methods involve time-consuming procedures, derivatization and/ or sophisticated instruments.

Due to the fact that MOX is a compound of great pharmacological and analytical importance, in recent years, there has been an increased interest to develop accurate analytical methods which are valid for quantification of MOX in biological and pharmaceutical samples.

Potentiometric methods, using ion selective electrodes, have found wide application [19]-[21] being simple, economical, applicable over a wide range of different areas, with applicability to turbid and colored solutions, and offering enough selectivity towards the drug in the presence of various pharmaceutical excipients. To the best of our knowledge till now no potentiometric membrane sensors for MOX have been published. The proposed sensors are based on the use of PVC membrane sensor of MOX - tetraphenylborate or MOX-phosphomolybdate or MOX-phosphotungstate as electroactive materials. The present work describes the construction and evaluation of novel PVC electrochemical sensors for the sensitive and selective determination of moxifloxacin in its pharmaceutical preparations.

Experimental

Apparatus

All potentiometric measurements were made at 25 ± 1°C unless otherwise stated using an Orion pH/mV meter (model 330) using MOX membrane sensors in conjunction with an Orion double junction Ag/AgCl reference electrode (model 90-02) containing 10% (w/v) potassium nitrate in the outer compartment. Adjustment of pH was made with a combined Ross glass pH electrode (Orion 81-02) for all pH measurements.

Reagents and materials

All chemicals used were of analytical reagent grade unless otherwise stated and doubly distilled water was used throughout. Polyvinyl chloride powder (PVC) high molecular weight, dibutyl sebacate (DBS), dioctyl phthalate (DOP), o-nitrophenyl octylether (NPOE), tetrahydrofurane (THF) of purity > 99% were obtained from Aldrich Chemical Company and MOX was obtained from Sigma Chemical Company, Germany. Sodium tetraphenylborate (NaTPB), phosphomolybdic acid (PMA) and phosphotungstic acid (PTA), were obtained from BDH, Chemical Ltd. Avelox 400 mg, Manufactured by Bayer was obtained from local pharmacy. The stock solution of 1 × 10-2 M MOX was prepared by dissolving the appropriate amount of MOX in 100 ml of water. Five standard MOX solutions were prepared in the range of 1 × 10-2 - 1 × 10-6 M by diluting the appreciate amount in double distilled water. Tris buffer of pH 7.0 was prepared by mixing 100 ml of 0.1 M tris(hydroxymethyl) aminomethane hydrochloride, with appropriate 0.1 M HCl.

Preparation of the MOX-PVC membrane sensors

Upon the addition of 25 ml of 1 × 10-2 M of MOX solution to 25 ml each of 1 × 10-2 M sodium tetraphenyl borate or 75 ml of 1 × 10-2 M of MOX solution to 25 ml phosphotungstic acid respectively, a whitish precipitate of MOX-TPB or yellowish precipitate of MOX-PA or MOX-PT were formed, respectively. The precipitate was filtered off through a Whatman filter paper No. 42, washed with cold deionized water until no chloride ion was detected into the washing solution. The precipitate was dried under vacuum for 48 h, then grinded to a fine powder in mortar, forming ion-pairs complex. The elemental analysis confirmed the formation of 1:1 or 3:1 complex of MOX:TPB or MOX:PM or MOX:PT, respectively. Portion of ten mg of the prepared ion associate complexes were thoroughly mixed with 190 mg PVC powder, 350 mg of DBS or DOP or NPOE and 5 ml THF in glass Petri dishes (5°Cm diameter). After the constituents being well mixed, the solvent has been allowed to evaporate overnight while the sensing membranes have been formed. The PVC master membranes were sectioned with a cork borer (10 Mm diameter) and glued to a polyethylene tube (3°Cm length, 8 Mm I.D.) using THF [22],[23]. Laboratory made electrode bodies were used, which consisted of a glass tube, to which the polyethylene tube is attached at one end and filled with internal reference solution (equal volumes of 1 × 10-2 M aqueous solution of MOX and KCl). Ag/AgCl internal reference electrode (1.0 Mm diameters) was used. The indicator electrode was conditioned by soaking in a 1 × 10-2 M aqueous MOX solution for 1 h and stored in the same solution when not in use.

Procedure

The moxifloxacin PVC membrane sensors were calibrated by immersion in conjunction with the reference electrode in a 50 ml beaker containing 9.0 ml of tris-buffer of pH 7.0. Then 1.0 ml aliquot of MOX solution was added with continuous stirring, to give final MOX concentration ranging from 1 × 10-2 to 1 × 10-6 M and the potential was recorded after stabilization to × 0.5 MV. A calibration curves were constructed by plotting the recorded potentials as a function of -log [MOX]. The resulting graphs were used for subsequent determination of unknown moxifloxacin concentration.

Determination of moxifloxacin in the pharmaceutical dosage forms

Ten tablets of Avelox 400 mg were accurately weighed, crushed, mixed in a mortar. An appropriate amount (400 mg of moxifloxacin powder, from each) was weighed, transferred to a 100 ml beaker and dissolved in double distilled water, sonication for about 15 Min and completed to the mark with the water. A 5.0 ml aliquots of these solutions were transferred to 50 ml standard flask, the pH was adjusted to 7.0 using tris buffer and completed to the mark with water. The potential of the solution was measured using MOX-sensors in conjunction with an Orion Ag/AgCl double junction reference electrode. The potential of the stirred solution was recorded after the signal stabilization (±0.5 MV/min) and the concentration was calculated from the previous calibration graph under identical experimental conditions from standard solutions of MOX.

Alternatively, the potentials displayed by moxifloxacin test solution before and after the addition of a 1.0 ml aliquot of 1 × 10-3 M moxifloxacin were measured. The change in the potential readings was recorded and used to calculate the unknown moxifloxacin concentration in the test solution using the standard addition technique [24].

Reconstituted powder: one mixture was prepared with a known amount of moxifloxacin powdered (20 mg) and other components such as starch, lactose and magnesium stearate. The accuracy of the potentiometric determination of MOX in this powdered was checked by evaluation the recovery.

Results and discussion

Sodium tetraphenyl borate, phosphomolybdic acid, and phosphotungstic acid were tested as ion-pairing agent for the preparation of electroactive ion association complexes for MOX. Sparingly soluble complexes of MOX-TPB, MOX- PM or MOX - PT have been instantaneously formed upon the addition of MOX solution to solutions of Na TPB, PMA or PTA respectively. The dry powder of the formed ion pairs are used for the construction of a new moxifloxacin ion selective electrodes. The elemental analysis showed that the composition of the complex is 1:1 in case of MOX:TPB, 3:1 for MOX: PM or MOX-PT respectively. Plasticized polymeric membranes were prepared by using membrane cocktails with compositions 1.82% of the corresponding ionic pair (MOX-TPB or MOX-PM or MOX-PTA), 34.54% of PVC and 63.64% of the corresponding plasticizer (DBS, DOP and NPOE).

Effect of plasticizer type on the characteristic performance of the sensors

Moxifloxacin ion-selective membrane sensors with different electroactive materials were investigated in order to compare their performance. Three reagents were investigated as possible counter ion for the preparation of the electroactive complex of MOX, namely TPB or PM or PTA were tested as ion-pair reagents. The obtained ion-pairs combined with three plasticizer, DOP, DBS and NPOE to give different combinations were tested. It is well known that the construction of PVC based ISEs required the use of a plasticizer which acts as a fluidizer allowing homogenous dissolution and diffusion mobility of the ion-pair inside the membrane. PVC membrane sensor of MOX-TPB, MOX-PM or MOX-PT with different plasticizer namely (DBS or DOP or NPOE) was found to be all suitable and optimum available mediators for MOX membrane sensors. In fact, o-NPOE was found to be the optimum available mediator for MOX-TPB, MOX-PM or MOX-PT membrane sensors (ion-associates). The use of non polar mediators such as DBS, DOP gave less solubility of the ion-pair and less response of the analyte compared with o-NPOE.(slope about 50.0 MV per concentration decade for both DBS and DOP respectively). It seems that o-NPOE improves the membrane selectivity due to its high dialectical constant (ε = 24), affects considerable dissolution of ion-association within the membrane; consequently enhances its partition coefficient in the membrane and also provided suitable mechanical property of the membrane compared with less permittivity plasticizers of DBS (ε = 4) or DOP (ε = 7) and the solubility of electroactive materials are relatively small compared with NPOE. Ortho-nitrophenyl octylether was used in case of MOX-TPB, MOX-PM or MOX-PT for carrying out other experiments in this investigation.

Effect of pH and the response time

The electrode response for different moxifloxacin concentrations was tested at different pH values, the pH being adjusted using hydrochloric acid or sodium hydroxide. The MOX-PVC electrode dipped into MOX solution of 1 × 10-3 and 1 × 10-4 M the potential of the electrode was plotted against the pH of solution. The potentials show that the slope per concentration decade is constant (~53.0 × 0.5 or 54.5 × 0.5 or 55.0 × 0.5 MV for MOX -PM or MOX -PT respectively) in the pH range of 6-9.0 (Figure 2). At higher pH values (>9.5), the potential decreased due to the gradual increase in the concentration of the unprotonated MOX.

Figure 2
figure 2

Effect of pH on the response of moxifloxacin sensors A) MOX-TPB, B) MOX-PM and C) MOX-PT using two series of moxifloxacin solutions: 1 × 10 -3 M, and 1 × 10 -4 M.

Average response time is defined [25] as the time required for the electrode to reach a stable potential within ±1 MV of the final equilibrium value. After successive immersion of the electrode in different moxifloxacin solutions each having a 10-fold difference in concentration or after rapid 10-fold increase in concentration by addition of MOX. The response time was found to be 20s for concentration of ≥1 × 10-3 M and ≥ 30 s for concentration 1 × 10-4 M. Day-to-day reproducibility of the sensor is about × 0.5 MV for the same solution and the useful lifetime of the sensor is 4 weeks, during which the potential slope is reproducible to within ±1 MV/ decade. Also after more than one month a new section from the master membrane was found to work with a good reproducibility.

Effect of diverse ions

The influences of different organic and inorganic ions on the response of MOX sensors were investigated. The selectivity coefficients K MF , B Pot were evaluated according to IUPAC guidelines using the separate solution method (SSM) or mixed solution method [25],[26] in tris-buffer solution of pH 7.0. The selectivity coefficient K MOX , B pot measured by separate solution method was calculated from the following equation:

log K A , B pot = E B - E A /S+ 1 - Z A / Z B log a A
(1)

where EA and EB are the potential reading observed after 1 Min of exposing the sensor to the same concentration of MOX and interfering species (1 × 10-3 each) alternatively. The symbol aA , and aB are the activity of MOX and interfering species and ZA and ZB are the charge of moxifloxacin and interfering species and S is slope of calibration graph (mV/ concentration). The selectivity coefficient by mixed solution method was defined as the activity ratio of primary and interfering ions that give the same potential change under identical conditions as given in equation 3.

K A , B pot = a ` A - a A / a B
(2)

Where a`A known activity of primary ion solution added into a reference solution that contains a fixed activity (aA) of primary ions, and the corresponding potential change (ΔE) is recorded. Next, a solution of an interfering ion (aB) is added to the reference solution until the same potential change (ΔE) is recorded. The change in potential produced at the constant back ground of the primary ion must be the same in both cases. The results are given in Table 1. The results reveal reasonable selectivity for MOX in presence of many related substances.

Table 1 Potentiometric selectivity coefficients of some interfering ions, using MOX

Sensors characteristics

The potentiometric response characteristics of the moxifloxacin sensors based on the use of MOX-TPB, MOX-PM or MOX-PT ion pair complexes and DBS or DOP or NPOE as a plasticizer in a PVC matrixes were evaluated according to IUPAC recommendations [25]. Results in Table 2 show the characteristics performance of the PVC membrane sensors. The least squares equations obtained from the calibration data as follows:

E mV =Slog MOX +Intercept
(3)

where E, is the potential of the electrode, S equal slope of the electrodes (53.0 × 0.5, 54.5 × 0.5 and 55.5 × 0.5 MV for MOX-TPB, MOX-PM, and MOX-PT, respectively) and intercept (137.8 × 0.5, 143.04 × 0.5 and 141.78 × 0.5 for MOX-TPB, MOX -PM, and MOX -PT, respectively).

Table 2 Response characteristics of moxifloxacin-PVC matrix membrane sensors

Validity of the proposed method

Limit of quantification and limit of detection

Each of different concentration of standard solution was tested five times. The potentials obtained for the five analyses were averaged at each concentration. The average potential was plotted versus concentration. The relation between potential and concentration is logarithmic (equation 1) X = S log [MOX] + Y, where X is equal the potential, S is the slope, and Y is the intercept and ( r ) is the correlation coefficient. The sensors display a linear response over the concentration range 1 × 102 to 4 × 10-6, 1 × 102 to 5.0 × 10-6, 1 × 102 to 5.0 × 10-6 M), respectively over a pH range of 6.0 - 9.0. The limits of detection (LOD) and limits of quantification (LOQ) were determined according the IUPAC recommendation [24]. The lower limit of detection (LOD) defined as the concentration of MOX corresponding to the intersection of the extrapolated linear segment of the calibration graph which is of 3 × 10-6, 4 × 10-6 and 4.0 × 10-6 M for sensor 1, 2 and 3 respectively( Figure 3).

Figure 3
figure 3

Calibration curve of MOX membrane sensors.

Recovery

The recoveries of MOX were calculated by comparing the potential of the measured or found concentration to direct added standard in tris buffer pH 7.0. The assay of recovery, at each concentration, was computed using the following equation:

Recovery % = found concentration / added concentration x100.
(4)

The average recovery of the direct determinations of 400 μg/ml of MOX was 98.5 , 99.1 and 98.6% with RSD value of 1.76, 1.65 and 1.78% for sensor 1 , 2 and 3 respectively (Table 3).

Table 3 Day to day reproducibility of the proposed membrane sensors

Precision and accuracy of the method

The intra-day, inter-day accuracy and precision of the assays were investigated [27] by the analysis of MOX at 400 μg/ml in five replicate over a period of three days. The five replicate were subject to estimate the intra-day and inter-day precision. Calibration curves were prepared and analyzed daily and linear models were used to determine concentrations in the quality control samples. Percent accuracy was determined (using the data from the precision assessment) as the closeness of found concentration to the added standards. Precision was reported as% RSD. The results obtained (Table 3) are within the acceptance range of less than 3.0% (precision) and more than 98.3% for the accuracy.

Ruggedness

The ruggedness of the potentiometric method was evaluated [27] by carrying out the analysis using two different analyst (operator) and different instruments on different days. The RSD of less than 3.0% were observed for repetitive measurements in three different day time periods using two different instruments and operators. The results indicate that the method is capable of producing results with high precision.

Robustness

The robustness of the method is demonstrated [27] by the versatility of the experimental factors that affecting the potential response (e.g. pH and response time).

Preliminary inspection of the results under these various conditions suggested that the method is fairly robust, but the pH of the measuring solution should be in the range of 6.0-9.0. The optimum pH 7.0 was used using tris buffer.

Determination of MOX in its pharmaceutical preparation

The applicability of the MOX membrane sensors for determination of the drug in the dosage forms was firstly checked by the studying the recovery of an accurate amount of pure MOX in solutions.

The analysis of 2.0 - 4000.0 μg/ml MOX solutions (in five replicate) by direct potentiometry gave an average recovery of 98.1, 98.25 and 98.51% with a relative standard deviation of 1.76, 1.65 and 1.78% for sensor 1, 2 and 3 respectively, results are shown in Table 4.

Table 4 Direct determinations of moxifloxacin using PVC membrane sensors

The applicability of the MOX-membrane sensors to the determination of the drug in the dosage forms was firstly checked by studying the recovery of an accurate amount of pure MOX in a reconstituted powder samples. The recovery obtained from five measurements was found to be 98.0 or 98.0 or 98.5% with a relative standard deviation of 1.7 or 1.6 or 1.7% for MOX- TPB or MOX-PM or MOX-PT respectively. On the other hand, the determination of MOX in its formulations show an average recovery of 98.0 or 98.5 or 98.5% with relative standard deviation of 1.7 or 1.6 or 1.7% for sensor 1 or 2 or 3, respectively, results are shown in Table 5.

Table 5 Determination of moxifloxacin in some pharmaceutical preparations using the proposed membrane sensors

Results obtained for the analysis of MOX in its formulation by direct measurements using the proposed sensors and the standard USP pharmacopeia method [28] are given in Table 5. The data obtained in Table 5 proves that the potentiometeric method shows a high degree of precision and accuracy compared with US pharmacopoeia method.

Application of MOX-PVC sensors as indictor electrodes

The developed electrodes in conjunction with an Ag/AgCl reference electrode have been examined as an end point indicator electrode for potentiometric titrations of the drug. Titration MOX with sodium tetraphenylborate using MOX-TPB or MOX-PM or MOX-PT sensors has been performed (Figure 4). From the results it is clear that MOX reacts with NaTPB in the molar ratio of 1:1. The titration curves were symmetrical with a very well defined potential jump of about 250 MV for MOX-TPB, MOX-PM and MOX-PT respectively, indicating the high sensitivity of the electrodes.

Figure 4
figure 4

Typical potentiometric titration curves of 5.0 ml of 1 × 10 2 M MOX with 1 × 10 2 M sodium tetraphenylborate using moxifloxacin membrane sensors.

Conclusion

Three different ion-pair complexes of MOX have been performed as sensors for MOX, the MOX membrane sensor displayed good analytical performance. The sensors display a fast, stable and near-Nernstian response over a relative wide moxifloxacin concentration range (1 × 102 - 4.0 × 10-6, 1 × 102 - 5.0 × 10-6, 1 × 102 - 5.0 × 10-6 M) for sensor 1, 2 and 3, respectively over a pH range of 6.0 - 9.0. The direct determination of moxifloxacin show an average recovery of 98.5, 99.1 and 98.6% and a mean relative standard deviation of 1.8, 1.6 and 1.8% at 400 μg/ml for sensors 1, 2 and 3 respectively. The results obtained are within the acceptance range of less than 3.0% (precision) and more than 98.3% for the accuracy. The sensors have been used as indicator electrodes for potentiometric titration of moxifloxacin.

Authors' contributions

MMH coordinated the study and modified the text, AMAH conducted the method development, MAA coordinated the study and modified the text, AMA coordinated the study and modified the text, AA coordinated the study and modified the text and GAEM proposed the subject, designed the study, participated in the results discussion and revised the manuscript. All authors read and approved the final manuscript.

References

  1. Sweetman SC: Martindale: The complete drug reference. 2009, The pharm. press, Lon

    Google Scholar 

  2. Maryadele JO: The Merck Index: An encyclopedia of chemicals, drugs and biologicals. 2001, Merck Research Laboratories, Division of Merck and Co.,Inc. Whitehouse station, New Jersey

    Google Scholar 

  3. Drlica K, Zhao X: DNA gyrase, topoisomerase IV, and the 4-quinolones. Microbiol Mol Biol Rev. 1997, 61 (3): PMID 9293187. PMC 232616, 377-92-

    Google Scholar 

  4. Motwani SK, Chopra S, Ahmad FJ, Khar RK: Validated spectrophotometric methods for the estimation of moxifloxacin in bulk and pharmaceutical formulations. Spectrochim Acta A. 2007, 68: 250-6. 10.1016/j.saa.2006.11.023.

    Article  Google Scholar 

  5. Elbashir AA, Ebraheem SA, Elwagee AH, Aboul-Enein HY: New spectrophotometric methods for the determination of moxifloxacin in pharmaceutical formulations. Acta Chim Slov. 2013, 60: 159-65.

    CAS  Google Scholar 

  6. Kamruzzaman M, Alam AM, Lee SH, Ragupathy D, Kim YH, Park SR, Kim SH: Spectrofluorimetric study of the interaction between europium(III) and moxifloxacin in micellar solution and its analytical application. Spectrochim Acta A. 2012, 86: 375-80. 10.1016/j.saa.2011.10.051.

    Article  CAS  Google Scholar 

  7. Al-Ghannam SM: Atomic absorption spectroscopic, conductometric and colorimetric methods fordetermination of some fluoroquinolone antibacterials using ammonium reineckate. Spectrochim Acta A. 2008, 69: 1188-94. 10.1016/j.saa.2007.06.023.

    Article  Google Scholar 

  8. Radi AE, Wahdan T, Anwar Z, Mostafa H: Electrochemical determination of gatifloxacin, moxifloxacin and sparfloxacin fluoroquinolonic antibiotics on glassy carbon electrode in pharmaceutical formulations. Drug Test Anal. 2010, 2: 397-400. 10.1002/dta.143.

    Article  CAS  Google Scholar 

  9. Xu YH, Li D, Liu XY, Li YZ, Lu J: High performance liquid chromatography assay with ultraviolet detection formoxifloxacin: validation and application to a pharmacokinetic study in Chinese volunteers. J Chromatogr B. 2010, 878: 3437-41. 10.1016/j.jchromb.2010.10.024.

    Article  CAS  Google Scholar 

  10. Lemoine T, Breilh D, Ducint D, Dubrez J, Jougon J, Velly JF, Saux MC: Determination of moxifloxacin (BAY 12-8039) in plasma and lung tissue by high-performance liquid chromatography with ultraviolet detection using a fully automated extraction method with a new polymeric cartridge. J Chromatogr B. 2000, 742: 247-54. 10.1016/S0378-4347(00)00166-3.

    Article  CAS  Google Scholar 

  11. Guerra FL, Paim CS, Steppe M, Schapoval EE: Biological assay and liquid chromatographic method for analysis of moxifloxacin in tablets. J AOAC Int. 2005, 88: 1086-92.

    CAS  Google Scholar 

  12. Kumar AK, Ramachandran G: Simple and rapid liquid chromatography method for determination of moxifloxacin in plasma. J Chromatogr B. 2009, 877: 1205-8. 10.1016/j.jchromb.2009.02.042.

    Article  Google Scholar 

  13. Kumar AK, Sudha V, Srinivasan R, Ramachandran G: Simple and rapid liquid chromatography method for determination of moxifloxacin in saliva. J Chromatogr B. 2011, 879: 3663-7. 10.1016/j.jchromb.2011.09.047.

    Article  CAS  Google Scholar 

  14. Goudah A: Pharmacokinetics and tissue residues of moxifloxacin in broiler chickens. Br Poult Sci. 2009, 50: 251-8. 10.1080/00071660802710108.

    Article  CAS  Google Scholar 

  15. Cruz LA, Hall R: Enantiomeric purity assay of moxifloxacin hydrochloride by capillaryelectrophoresis. J Pharm Biomed Anal. 2005, 38: 8-13. 10.1016/j.jpba.2004.12.004.

    Article  CAS  Google Scholar 

  16. Möller JG, Stass H, Heinig R, Blaschke G: Capillary electrophoresis with laser-induced fluorescence: a routine method to determine moxifloxacin in human body fluids in very small sample volumes. J Chromatogr B. 1998, 716: 325-34. 10.1016/S0378-4347(98)00302-8.

    Article  Google Scholar 

  17. Raju B, Ramesh M, Borkar RM, Padiya R, Banerjee SK, Srinivas R: Development and validation of liquid chromatography-mass spectrometric method for simultaneous determination of moxifloxacin and ketorolac in rat plasma: application to pharmacokinetic study. Biomed Chromatogr. 2012, 26: 1341-7. 10.1002/bmc.2701.

    Article  CAS  Google Scholar 

  18. Pranger AD, Alffenaar JW, Wessels AM, Greijdanus B, Uges DR: Determination of moxifloxacin in human plasma, plasma ultrafiltrate, and cerebrospinal fluid by a rapid and simple liquid chromatography- tandem mass spectrometry method. J Anal Toxicol. 2010, 34: 135-41. 10.1093/jat/34.3.135.

    Article  CAS  Google Scholar 

  19. Elzanfaly ES, Zaazaa HE, Merey HA: Ion selective phosphotungestate and beta-cyclodextrin based membraneelectrodes for stability-indicating determination of midodrine hydrochloride. Acta Chim Slov. 2013, 60: 256-6.

    CAS  Google Scholar 

  20. Hassan SS, Kamel AH, Abd El-Naby H: New potentiometric sensors based on selective recognition sites for determination of ephedrine in some pharmaceuticals and biological fluids. Talanta. 2013, 103: 330-6. 10.1016/j.talanta.2012.10.067.

    Article  CAS  Google Scholar 

  21. Abu Shawish HM, Khedr AM, Abed-Almonem KI, Gaber M: A comparative study of solid and liquid inner contact benzalkonium chloride ion-selective electrode membranes. Talanta. 2012, 101: 211-9. 10.1016/j.talanta.2012.09.005.

    Article  CAS  Google Scholar 

  22. Hassan SSM, Marzouk SAM: A novel ferroin membrane sensor for potentiometric determination of iron. Talanta. 1994, 41: 891-899. 10.1016/0039-9140(94)E0042-P.

    Article  CAS  Google Scholar 

  23. Carggs A, Moody GJ, Tomas JDR: PVC matrix membrane ion-selective electrodes. J Chem Educ. 1974, 51: 541-544. 10.1021/ed051p541.

    Article  Google Scholar 

  24. Ma TS, Hassan SSM: Organic Analysis Using Ion Selective Electrodes. 1982, Academic Press, London

    Google Scholar 

  25. Analytical Chemistry Division IUPAC: Recommendation for Nomenclature of Ion Selective Electrode. Pure Appl Chem. 1994, 66: 2527-2536.

    Google Scholar 

  26. Analytical Chemistry Division IUPAC: Potentiometric selectivity coefficients of ion selective electrodes. Pure Appl Chem. 2000, 72: 1851-2082.

    Google Scholar 

  27. Miller JC, Miller JN: Statistics for Analytical Chemistry. 1986, Ellis Harwood Limited, England

    Google Scholar 

  28. Non-US Monographs Guideline and is not a USP-NF monograph, moxifloxacin Tablets. 2011, The United States Pharmacopeia

Download references

Acknowledgements

The authors extend their appreciation to the Deanship of Scientific Research at King Saud University for funding the work through the research group project No. RGP-VPP-037.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Gamal A Mostafa.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ original submitted files for images

Rights and permissions

Open Access  This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hefnawy, M.M., Homoda, A.M., Abounassif, M.A. et al. Potentiometric determination of moxifloxacin in some pharmaceutical formulation using PVC membrane sensors. Chemistry Central Journal 8, 59 (2014). https://doi.org/10.1186/s13065-014-0059-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13065-014-0059-y

Keywords